43
9500+ SCPI Command Summary
Keyword
Lower-Level
Keyword
Lower-Level
Keyword
Parameter
Std /
New
Comments
:SPULSe Std
Subsystem. Contains commands to control
the output pulse generation. Commands
without suffix refer to the currently selected
logical instrument. See INSTrument
:STATe 0 / 1 or OFF / ON Std
Enables / Disables the output for all
channels. Command is the same as pressing
the RUN / STOP button.
:PERiod 0.0000002s - 5000s Std
Sets the T
0
period.
:MODe
NORMal / SINGle /
BURSt / DCYCle
New
Sets the T
0
mode.
:BCOunter 1 - 1000000 New
Burst Counter. Sets the number of pulses to
generate when in burst mode.
:PCOunter 1 - 1000000 New
Pulse Counter. Sets the number of pulses to
generate during the on cycle of the Duty
Cycle mode.
:OCOunter 1 - 1000000 New
Off Counter. Number of pulses to inhibit
output during the off cycle of the Duty Cycle
mode.
:EXTernal New
Subsystem. Contains the commands to
define the system use of the external input.
:MODe
DISabled / TRIGger
/ GATe
New
Selects the trigger mode.
:LEVel 0.20 - 15 New
Sets the trigger threshold. Value is in volts,
with a range of 0.20 to 15 Volts.
:EDGe RISing / FALLing New
Selects which edge (rising of falling) to use
as the trigger signal.
:POLarity LOW / HIGH New
Sets the POLarity of the gate signal. HIGH
output is active when the gate signal is high.
LOW output is active when the gate signal is
Comentarios a estos manuales