44
9500+ SCPI Command Summary
Keyword
Lower-Level
Keyword
Lower-Level
Keyword
Parameter
Std /
New
Comments
:PULSe [1 / 2 / n]
Std
Subsystem. Contains commands to
control the output pulse generation. Valid
suffix range depends on the number of
channels (ChA = 1, ChB = 2, ect).
Command without suffix refers to the
currently selected logical instrument. See
INSTrument subsystem.
:STATe
0 / 1 or OFF / ON
Std
Enables / Disables the output pulse for the
selected channel.
:WIDTh
Std
Sets the width or duration of the output
pulse.
:DELay
-1000s - 1000s
Std
Sets the time from the start of the T
0
period to the first edge of the pulse.
:SYNC
T0, CHA, CHB,
CHC, CHD, ect.
New
Selects the Sync Source.
:MUX
0 - 255
New
Selects which tmiers are enabled as
outputs for the current channel. The input
is a decimal representation of an 8 bit
binary number (example: 255 = 1111
:POLarity
NORMal /
COMPlement /
INVerted
Std
Sets the polarity of the pulse. For
NORMal operation, the second nominal
state is more positive that the first.
COMPlement and INVerted are aliases.
For both, the second state is more negative
than the first.
:OUTPut
New
Subsystem. Contains commands to
control the output mode.
:AMPLitude
2.00 - 20.00
New
Sets adjustable output level.
:MODe
/ 35V
New
Selects output AMPLitude mode:
TTL/CMOS, ADJustable, or 35 Volts.
:CMODe
SINGle / BURSt /
DCYCle
New
Channel Mode. Sets the channel pulse
series output mode.
:BCOunter 1 - 1000000 New
Burst Counter. Sets the number of pulses
to generate when the channel is in the
BURST mode.
:PCOunter 1 - 1000000 New
Pulse Counter. Sets the number of pulses
to generate during the on cycle of the Duty
Cycle mode.
:OCOunter 1 - 1000000 New
Off Counter. Number of pulses to inhibit
output during the off cycle of the Duty
Cycle mode.
:CGATe
DISable / LOW /
HIGH
New
Channel Gate Subsystem. Contains
commands to control using the gate input
to control the output channel.
:WCOunter
1 - 1000000
New
0
pulses to delay until
enabling the output.
NOTE: Output Amplitude is limited to 20V for both standard 9500+and 8510 units.
NOTE: 35V output mode only applicable on 9500+ 35V option.